*** MIDAS logfile (05SEPpl1.2) *** Su, 13 May 2007 16:59:23 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Su, 13 May 2007 16:59:37 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 070513 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 356.53 stddev = 1.96 FIERA temperature = 273.15 K There were 2 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_070513 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 2.17 1669.5 1669.7 797.0 0.17 0.14 2.09 2 4.34 3325.9 3326.5 1585.1 -0.23 -0.23 2.10 3 8.67 6661.3 6660.8 3225.7 -0.08 -0.09 2.07 4 13.00 9988.1 9988.2 4852.7 -0.13 -0.13 2.06 5 19.51 14988.8 14989.7 7225.8 -0.09 -0.08 2.07 6 26.01 19996.6 19994.0 9817.9 -0.03 -0.03 2.04 7 32.51 25006.1 25006.4 12022.6 0.01 0.02 2.08 8 43.34 33367.6 33366.4 15976.4 0.09 0.10 2.09 9 54.17 41729.9 41731.6 19693.9 0.14 0.15 2.12 10 65.01 50074.5 50071.9 23344.1 0.14 0.14 2.15 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3564 to 106900 e- Average count rate measured: 769 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.13 +/- 0.02 e-/ADU RON = 4.18 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.004 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.16/-0.23 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay