*** MIDAS logfile (05SEPpl1.2) *** We, 02 May 2007 15:05:55 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** We, 02 May 2007 15:06:10 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 070502 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 356.13 stddev = 1.97 FIERA temperature = 273.15 K There were 2 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_070502 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 2.17 1684.1 1684.1 812.9 0.15 0.17 2.07 2 4.34 3354.3 3353.9 1618.8 -0.24 -0.25 2.07 3 8.67 6718.6 6718.1 3243.7 -0.08 -0.08 2.07 4 13.00 10073.0 10072.0 4843.4 -0.14 -0.14 2.08 5 19.51 15118.5 15117.4 7282.7 -0.08 -0.08 2.08 6 26.01 20167.1 20166.0 9704.4 -0.03 -0.03 2.08 7 32.51 25224.9 25222.9 12167.5 0.03 0.03 2.07 8 43.34 33650.2 33648.0 16077.2 0.09 0.08 2.09 9 54.17 42099.1 42100.1 19845.6 0.17 0.18 2.12 10 65.01 50494.5 50490.5 23910.6 0.12 0.12 2.11 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3564 to 106843 e- Average count rate measured: 776 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.12 +/- 0.01 e-/ADU RON = 4.17 +/- 0.02 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.003 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.18/-0.25 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay