*** MIDAS logfile (05SEPpl1.2) *** Sa, 28 Apr 2007 15:52:06 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Sa, 28 Apr 2007 15:52:21 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 070428 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 355.93 stddev = 1.97 FIERA temperature = 273.15 K There were 2 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_070428 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 2.17 1686.3 1685.4 817.2 0.17 0.14 2.06 2 4.34 3359.7 3359.9 1615.8 -0.22 -0.21 2.08 3 8.67 6727.3 6726.4 3178.9 -0.10 -0.11 2.12 4 13.00 10088.8 10088.3 4909.2 -0.13 -0.13 2.06 5 19.51 15140.1 15141.4 7248.2 -0.08 -0.07 2.09 6 26.01 20197.4 20195.9 9746.9 -0.03 -0.03 2.07 7 32.51 25257.4 25258.0 12099.9 0.01 0.02 2.09 8 43.34 33704.5 33706.7 16260.7 0.09 0.10 2.07 9 54.17 42147.1 42147.8 19855.7 0.13 0.14 2.12 10 65.01 50584.8 50585.3 23391.4 0.15 0.15 2.16 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3612 to 108366 e- Average count rate measured: 777 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.14 +/- 0.02 e-/ADU RON = 4.23 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.004 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.16/-0.21 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay