*** MIDAS logfile (05SEPpl1.2) *** Mo, 23 Apr 2007 16:01:05 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Mo, 23 Apr 2007 16:01:19 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 070423 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 355.88 stddev = 1.98 FIERA temperature = 273.15 K There were 2 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_070423 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 2.17 1688.6 1688.9 813.5 0.13 0.16 2.08 2 4.34 3367.3 3366.2 1597.9 -0.19 -0.20 2.11 3 8.67 6740.9 6740.2 3146.5 -0.10 -0.10 2.14 4 13.00 10108.8 10108.3 4852.0 -0.13 -0.13 2.08 5 19.51 15170.2 15170.0 7366.9 -0.08 -0.09 2.06 6 26.01 20238.1 20236.8 9926.0 -0.03 -0.04 2.04 7 32.51 25309.0 25308.7 12133.6 0.02 0.01 2.09 8 43.34 33773.4 33774.1 16225.2 0.10 0.10 2.08 9 54.17 42232.1 42232.9 20140.3 0.14 0.13 2.10 10 65.01 50687.2 50687.1 23444.9 0.15 0.15 2.16 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3601 to 108078 e- Average count rate measured: 779 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.13 +/- 0.02 e-/ADU RON = 4.22 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.005 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.16/-0.20 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay