*** MIDAS logfile (05SEPpl1.2) *** We, 11 Apr 2007 14:03:54 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** We, 11 Apr 2007 14:04:08 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 070411 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 356.14 stddev = 1.97 FIERA temperature = 273.15 K There were 2 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_070411 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 2.17 1702.8 1703.3 827.1 0.14 0.16 2.06 2 4.34 3394.3 3394.5 1591.9 -0.21 -0.20 2.13 3 8.67 6796.9 6795.8 3274.6 -0.09 -0.10 2.08 4 13.00 10191.4 10192.6 4934.2 -0.13 -0.12 2.07 5 19.51 15295.1 15294.6 7344.3 -0.08 -0.09 2.08 6 26.01 20402.6 20403.0 9822.9 -0.03 -0.04 2.08 7 32.51 25516.8 25516.5 12220.9 0.02 0.01 2.09 8 43.34 34049.8 34047.7 16466.6 0.10 0.09 2.07 9 54.17 42579.3 42581.1 20015.4 0.14 0.14 2.13 10 65.01 51098.8 51101.1 23777.3 0.14 0.15 2.15 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3637 to 109113 e- Average count rate measured: 785 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.14 +/- 0.02 e-/ADU RON = 4.22 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.004 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.15/-0.20 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay