*** MIDAS logfile (05SEPpl1.2) *** Sa, 31 Mar 2007 15:25:57 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Sa, 31 Mar 2007 15:26:12 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 070331 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 356.38 stddev = 1.97 FIERA temperature = 273.15 K There were 2 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_070331 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 2.17 1709.4 1709.7 840.0 0.14 0.15 2.03 2 4.34 3407.7 3408.1 1623.7 -0.21 -0.20 2.10 3 8.67 6823.6 6822.8 3269.9 -0.09 -0.10 2.09 4 13.00 10233.7 10232.0 4954.0 -0.12 -0.13 2.07 5 19.51 15355.6 15356.2 7274.3 -0.09 -0.08 2.11 6 26.01 20485.9 20485.3 9813.3 -0.03 -0.03 2.09 7 32.51 25618.5 25619.8 12455.9 0.01 0.02 2.06 8 43.34 34187.2 34185.3 15918.3 0.10 0.09 2.15 9 54.17 42748.1 42749.2 20215.0 0.13 0.13 2.11 10 65.01 51306.6 51307.2 24028.5 0.15 0.15 2.14 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3649 to 109527 e- Average count rate measured: 788 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.13 +/- 0.01 e-/ADU RON = 4.20 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.004 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.15/-0.21 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay