*** MIDAS logfile (05SEPpl1.2) *** Mo, 19 Mar 2007 14:12:20 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Mo, 19 Mar 2007 14:12:34 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 070319 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 355.61 stddev = 1.97 FIERA temperature = 273.15 K There were 2 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_070319 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 2.17 1705.2 1705.1 818.0 0.13 0.13 2.08 2 4.34 3400.7 3401.0 1618.1 -0.17 -0.17 2.10 3 8.67 6806.0 6806.3 3227.6 -0.10 -0.10 2.11 4 13.00 10207.7 10207.8 4925.5 -0.12 -0.12 2.07 5 19.51 15319.4 15317.7 7330.1 -0.07 -0.08 2.09 6 26.01 20433.2 20433.6 9754.7 -0.03 -0.04 2.09 7 32.51 25553.0 25554.6 12346.2 0.01 0.01 2.07 8 43.34 34101.2 34101.6 16415.7 0.10 0.10 2.08 9 54.17 42634.7 42637.7 20152.4 0.12 0.12 2.12 10 65.01 51177.9 51181.2 23997.2 0.15 0.15 2.13 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3617 to 108547 e- Average count rate measured: 786 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.12 +/- 0.01 e-/ADU RON = 4.17 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.004 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.15/-0.17 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay