*** MIDAS logfile (05SEPpl1.2) *** Th, 08 Mar 2007 16:15:49 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Th, 08 Mar 2007 16:16:03 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 070308 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 355.47 stddev = 1.98 FIERA temperature = 273.15 K There were 2 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_070308 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 2.17 1709.9 1710.3 818.8 0.13 0.13 2.09 2 4.34 3410.3 3409.4 1614.0 -0.18 -0.20 2.11 3 8.67 6825.0 6824.5 3270.0 -0.11 -0.12 2.09 4 13.00 10236.6 10237.2 4893.1 -0.12 -0.12 2.09 5 19.51 15361.7 15361.4 7486.6 -0.08 -0.08 2.05 6 26.01 20491.6 20492.7 9884.9 -0.03 -0.03 2.07 7 32.51 25624.2 25626.2 12375.3 0.00 0.01 2.07 8 43.34 34201.2 34202.1 16305.1 0.11 0.11 2.10 9 54.17 42757.7 42759.7 20181.7 0.12 0.13 2.12 10 65.01 51333.2 51330.6 23926.1 0.17 0.16 2.15 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3650 to 109550 e- Average count rate measured: 788 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.13 +/- 0.01 e-/ADU RON = 4.22 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.004 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.17/-0.19 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay