*** MIDAS logfile (05SEPpl1.2) *** Fr, 16 Feb 2007 16:29:27 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Fr, 16 Feb 2007 16:29:41 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 070216 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 355.32 stddev = 1.98 FIERA temperature = 273.15 K There were 2 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_070216 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 2.17 1717.0 1716.2 823.2 0.15 0.13 2.09 2 4.34 3422.6 3421.9 1618.0 -0.19 -0.20 2.12 3 8.67 6850.0 6849.4 3289.5 -0.11 -0.11 2.08 4 13.00 10274.2 10275.0 5060.7 -0.13 -0.11 2.03 5 19.51 15418.0 15418.0 7419.4 -0.08 -0.08 2.08 6 26.01 20567.2 20566.6 9832.9 -0.03 -0.03 2.09 7 32.51 25720.5 25720.7 12408.7 0.01 0.01 2.07 8 43.34 34325.5 34326.8 16620.9 0.10 0.11 2.07 9 54.17 42912.6 42915.1 20157.5 0.12 0.12 2.13 10 65.01 51515.1 51519.1 24007.3 0.15 0.16 2.15 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3664 to 109954 e- Average count rate measured: 791 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.13 +/- 0.02 e-/ADU RON = 4.22 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.004 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.16/-0.19 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay