*** MIDAS logfile (05SEPpl1.2) *** Tu, 13 Feb 2007 13:43:30 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Tu, 13 Feb 2007 13:43:44 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 070213 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 355.26 stddev = 1.97 FIERA temperature = 273.15 K There were 2 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_070213 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 2.17 1718.9 1719.8 817.7 0.15 0.15 2.10 2 4.34 3427.4 3426.5 1645.5 -0.18 -0.21 2.08 3 8.67 6858.9 6859.6 3269.1 -0.11 -0.10 2.10 4 13.00 10288.1 10287.8 4938.0 -0.12 -0.13 2.08 5 19.51 15438.6 15437.2 7530.9 -0.08 -0.09 2.05 6 26.01 20593.9 20594.4 9996.5 -0.04 -0.03 2.06 7 32.51 25755.0 25756.7 12487.9 0.01 0.02 2.06 8 43.34 34368.9 34369.4 16363.5 0.10 0.10 2.10 9 54.17 42976.3 42977.3 20255.8 0.13 0.14 2.12 10 65.01 51579.1 51580.9 24058.9 0.15 0.15 2.14 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3671 to 110127 e- Average count rate measured: 792 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.14 +/- 0.02 e-/ADU RON = 4.22 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.004 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.15/-0.20 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay