*** MIDAS logfile (05SEPpl1.2) *** Su, 04 Feb 2007 17:25:09 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Su, 04 Feb 2007 17:25:23 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 070204 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 355.53 stddev = 1.98 FIERA temperature = 273.15 K There were 2 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_070204 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 2.17 1705.6 1705.2 824.0 0.16 0.14 2.07 2 4.34 3400.1 3401.0 1625.7 -0.20 -0.17 2.09 3 8.67 6807.2 6806.7 3304.7 -0.10 -0.12 2.06 4 13.00 10210.8 10210.4 4929.5 -0.11 -0.12 2.07 5 19.51 15321.4 15322.3 7403.8 -0.09 -0.09 2.07 6 26.01 20439.5 20441.4 9790.5 -0.03 -0.03 2.09 7 32.51 25559.7 25561.4 12213.2 0.00 0.01 2.09 8 43.34 34113.4 34113.6 16313.5 0.10 0.10 2.09 9 54.17 42645.2 42649.1 20194.4 0.11 0.12 2.11 10 65.01 51197.5 51199.3 23881.2 0.16 0.16 2.14 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3637 to 109190 e- Average count rate measured: 786 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.13 +/- 0.01 e-/ADU RON = 4.23 +/- 0.02 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.005 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.16/-0.19 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay