*** MIDAS logfile (05SEPpl1.2) *** Th, 21 Dec 2006 16:31:06 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Th, 21 Dec 2006 16:31:21 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 061221 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 356.54 stddev = 1.96 FIERA temperature = 273.15 K There were 2 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_061221 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 2.17 1720.6 1720.4 826.9 0.15 0.13 2.08 2 4.34 3431.8 3432.4 1636.6 -0.17 -0.16 2.10 3 8.67 6867.7 6868.4 3300.6 -0.12 -0.11 2.08 4 13.00 10303.9 10302.8 5037.7 -0.11 -0.12 2.05 5 19.51 15460.4 15459.2 7436.8 -0.08 -0.09 2.08 6 26.01 20623.7 20626.2 9919.0 -0.03 -0.02 2.08 7 32.51 25788.5 25789.3 12469.4 -0.00 -0.00 2.07 8 43.34 34421.7 34420.3 16251.5 0.10 0.10 2.12 9 54.17 43025.1 43029.2 20376.7 0.10 0.11 2.11 10 65.01 51662.0 51663.9 23875.8 0.16 0.16 2.16 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3695 to 110953 e- Average count rate measured: 793 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.15 +/- 0.02 e-/ADU RON = 4.21 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.005 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.16/-0.16 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay