*** MIDAS logfile (05SEPpl1.2) *** We, 06 Dec 2006 15:44:07 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** We, 06 Dec 2006 15:44:21 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 061206 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 354.96 stddev = 2.02 FIERA temperature = 273.15 K There were 2 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_061206 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 2.17 1708.9 1708.7 822.6 0.13 0.15 2.08 2 4.34 3408.1 3407.8 1604.8 -0.18 -0.17 2.12 3 8.67 6821.0 6821.3 3296.8 -0.11 -0.10 2.07 4 13.00 10230.7 10230.3 4915.1 -0.12 -0.13 2.08 5 19.51 15352.0 15352.9 7436.2 -0.09 -0.09 2.06 6 26.01 20479.9 20480.4 9796.9 -0.03 -0.04 2.09 7 32.51 25610.5 25611.3 12252.3 0.00 -0.00 2.09 8 43.34 34182.3 34185.0 16596.4 0.11 0.11 2.06 9 54.17 42731.7 42733.3 20127.8 0.12 0.11 2.12 10 65.01 51304.9 51307.3 23907.9 0.17 0.16 2.15 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3641 to 109317 e- Average count rate measured: 788 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.13 +/- 0.02 e-/ADU RON = 4.30 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.005 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.17/-0.18 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay