*** MIDAS logfile (05SEPpl1.2) *** We, 29 Nov 2006 15:19:30 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** We, 29 Nov 2006 15:19:45 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 061129 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 355.19 stddev = 2.01 FIERA temperature = 273.15 K There were 2 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_061129 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 2.17 1712.7 1712.8 826.1 0.13 0.13 2.07 2 4.34 3415.1 3415.9 1609.7 -0.19 -0.18 2.12 3 8.67 6836.3 6836.6 3283.4 -0.10 -0.10 2.08 4 13.00 10253.8 10253.1 4948.7 -0.11 -0.12 2.07 5 19.51 15386.6 15386.9 7466.9 -0.08 -0.08 2.06 6 26.01 20524.0 20524.9 10037.2 -0.04 -0.03 2.04 7 32.51 25667.2 25667.8 12273.8 0.01 0.01 2.09 8 43.34 34255.4 34253.8 16287.7 0.10 0.10 2.10 9 54.17 42827.6 42830.6 20496.9 0.12 0.13 2.09 10 65.01 51405.9 51406.2 23967.9 0.15 0.15 2.14 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3641 to 109293 e- Average count rate measured: 790 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.13 +/- 0.02 e-/ADU RON = 4.27 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.004 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.15/-0.19 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay