*** MIDAS logfile (05SEPpl1.2) *** Su, 26 Nov 2006 13:23:30 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Su, 26 Nov 2006 13:23:45 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 061126 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 355.12 stddev = 2.01 FIERA temperature = 273.15 K There were 2 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_061126 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 2.17 1715.0 1714.8 831.7 0.13 0.12 2.06 2 4.34 3419.5 3419.7 1628.2 -0.18 -0.17 2.10 3 8.67 6843.7 6843.3 3255.7 -0.10 -0.11 2.10 4 13.00 10264.0 10264.3 4983.5 -0.12 -0.12 2.06 5 19.51 15402.0 15402.8 7501.2 -0.09 -0.08 2.05 6 26.01 20547.5 20547.6 9935.8 -0.03 -0.03 2.07 7 32.51 25695.3 25693.2 12483.8 0.01 0.00 2.06 8 43.34 34293.3 34293.2 16229.0 0.10 0.11 2.11 9 54.17 42873.7 42873.1 20258.7 0.12 0.12 2.12 10 65.01 51469.2 51469.6 23881.0 0.16 0.16 2.16 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3674 to 110283 e- Average count rate measured: 790 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.14 +/- 0.02 e-/ADU RON = 4.30 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.004 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.16/-0.18 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay