*** MIDAS logfile (05SEPpl1.2) *** We, 15 Nov 2006 14:35:07 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** We, 15 Nov 2006 14:35:22 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 061115 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 354.57 stddev = 2.01 FIERA temperature = 273.15 K There were 2 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_061115 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 2.17 1717.1 1716.2 818.9 0.14 0.15 2.10 2 4.34 3421.3 3421.8 1602.5 -0.21 -0.19 2.13 3 8.67 6850.1 6849.3 3329.6 -0.09 -0.10 2.06 4 13.00 10273.2 10272.1 4972.2 -0.11 -0.12 2.07 5 19.51 15414.1 15414.7 7445.3 -0.08 -0.08 2.07 6 26.01 20561.8 20561.6 10095.9 -0.03 -0.04 2.04 7 32.51 25716.5 25714.5 12536.8 0.02 0.01 2.05 8 43.34 34314.0 34313.4 16550.6 0.10 0.09 2.07 9 54.17 42906.7 42909.0 20171.9 0.13 0.13 2.13 10 65.01 51493.2 51495.5 23866.6 0.14 0.14 2.16 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3674 to 110211 e- Average count rate measured: 791 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.14 +/- 0.02 e-/ADU RON = 4.30 +/- 0.05 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.004 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.14/-0.20 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay