*** MIDAS logfile (05SEPpl1.2) *** Sa, 04 Nov 2006 12:56:32 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Sa, 04 Nov 2006 12:56:46 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 061104 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 355.45 stddev = 2.01 FIERA temperature = 273.15 K There were 2 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_061104 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 2.17 1711.8 1712.5 827.3 0.15 0.14 2.07 2 4.34 3412.4 3412.6 1597.8 -0.20 -0.21 2.14 3 8.67 6832.0 6831.3 3292.2 -0.09 -0.10 2.08 4 13.00 10244.9 10244.2 4983.4 -0.12 -0.13 2.06 5 19.51 15375.0 15374.7 7369.9 -0.08 -0.08 2.09 6 26.01 20509.5 20509.2 9936.3 -0.03 -0.03 2.06 7 32.51 25649.2 25650.2 12352.9 0.01 0.02 2.08 8 43.34 34224.1 34223.9 16153.1 0.09 0.09 2.12 9 54.17 42801.8 42800.2 20168.1 0.14 0.14 2.12 10 65.01 51362.0 51362.6 23588.3 0.14 0.15 2.18 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3696 to 110875 e- Average count rate measured: 789 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.16 +/- 0.02 e-/ADU RON = 4.33 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.004 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.15/-0.21 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay