*** MIDAS logfile (05SEPpl1.2) *** We, 25 Oct 2006 15:48:22 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** We, 25 Oct 2006 15:48:36 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 061025 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 355.36 stddev = 2.00 FIERA temperature = 273.15 K There were 2 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_061025 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 2.17 1716.6 1716.5 830.1 0.16 0.12 2.07 2 4.34 3420.6 3421.7 1623.1 -0.20 -0.20 2.11 3 8.67 6849.0 6848.7 3313.1 -0.09 -0.10 2.07 4 13.00 10270.3 10270.5 4971.7 -0.13 -0.13 2.07 5 19.51 15412.5 15412.8 7510.7 -0.08 -0.08 2.05 6 26.01 20561.2 20558.3 9953.6 -0.03 -0.04 2.07 7 32.51 25711.5 25712.4 12389.5 0.01 0.02 2.08 8 43.34 34313.0 34311.3 16392.8 0.10 0.10 2.09 9 54.17 42902.6 42905.2 20316.7 0.13 0.14 2.11 10 65.01 51493.6 51498.2 24400.8 0.15 0.16 2.11 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3627 to 108806 e- Average count rate measured: 791 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.11 +/- 0.01 e-/ADU RON = 4.22 +/- 0.02 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.004 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.16/-0.20 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay