*** MIDAS logfile (05SEPpl1.2) *** We, 18 Oct 2006 13:07:19 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** We, 18 Oct 2006 13:07:34 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 061018 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 356.08 stddev = 1.98 FIERA temperature = 273.15 K There were 2 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_061018 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 2.17 1703.2 1703.2 830.5 0.15 0.15 2.05 2 4.34 3395.9 3396.8 1607.5 -0.20 -0.18 2.11 3 8.67 6799.1 6799.4 3260.5 -0.10 -0.10 2.09 4 13.00 10196.3 10197.2 4924.6 -0.13 -0.13 2.07 5 19.51 15302.8 15302.4 7443.6 -0.09 -0.09 2.06 6 26.01 20414.9 20414.3 9868.1 -0.03 -0.03 2.07 7 32.51 25530.4 25529.7 12436.2 0.02 0.01 2.05 8 43.34 34069.5 34068.2 16476.3 0.10 0.09 2.07 9 54.17 42597.8 42600.2 20211.5 0.12 0.13 2.11 10 65.01 51129.4 51130.6 23950.4 0.15 0.15 2.13 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3610 to 108362 e- Average count rate measured: 785 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.12 +/- 0.02 e-/ADU RON = 4.19 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.005 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.15/-0.19 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay