*** MIDAS logfile (05SEPpl1.2) *** Fr, 13 Oct 2006 17:30:26 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Fr, 13 Oct 2006 17:30:41 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 061013 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 355.65 stddev = 1.98 FIERA temperature = 273.15 K There were 2 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_061013 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 2.17 1716.3 1716.6 830.2 0.14 0.14 2.07 2 4.34 3424.1 3422.9 1639.5 -0.16 -0.19 2.09 3 8.67 6851.1 6851.7 3288.5 -0.12 -0.12 2.08 4 13.00 10278.8 10278.6 4893.6 -0.11 -0.12 2.10 5 19.50 15422.3 15422.7 7467.4 -0.09 -0.09 2.07 6 26.01 20574.7 20576.6 9915.3 -0.03 -0.03 2.08 7 32.51 25727.8 25729.9 12472.2 0.00 0.01 2.06 8 43.34 34341.8 34342.8 16278.1 0.11 0.11 2.11 9 54.17 42922.9 42926.0 20565.3 0.10 0.11 2.09 10 65.01 51541.1 51545.0 23838.8 0.17 0.17 2.16 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3663 to 110011 e- Average count rate measured: 792 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.13 +/- 0.02 e-/ADU RON = 4.24 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.005 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.17/-0.17 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay