*** MIDAS logfile (05SEPpl1.2) *** Th, 12 Oct 2006 17:35:59 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Th, 12 Oct 2006 17:36:13 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 061012 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 355.39 stddev = 2.00 FIERA temperature = 273.15 K There were 2 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_061012 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 2.17 1706.2 1706.9 817.6 0.14 0.12 2.09 2 4.34 3403.4 3404.1 1639.1 -0.17 -0.17 2.08 3 8.67 6811.1 6810.5 3255.7 -0.11 -0.13 2.09 4 13.00 10216.7 10217.5 4908.7 -0.12 -0.11 2.08 5 19.51 15331.3 15332.3 7363.1 -0.09 -0.08 2.08 6 26.01 20452.9 20452.9 9893.3 -0.03 -0.03 2.07 7 32.51 25574.8 25576.0 12375.0 0.00 0.01 2.07 8 43.34 34133.8 34134.5 16148.3 0.10 0.11 2.11 9 54.17 42667.3 42672.1 20215.6 0.10 0.12 2.11 10 65.01 51233.2 51234.8 24112.9 0.16 0.17 2.12 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3621 to 108714 e- Average count rate measured: 787 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.12 +/- 0.01 e-/ADU RON = 4.24 +/- 0.02 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.005 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.17/-0.17 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay