*** MIDAS logfile (05SEPpl1.2) *** We, 04 Oct 2006 15:22:13 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** We, 04 Oct 2006 15:22:28 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 061004 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 356.06 stddev = 1.99 FIERA temperature = 273.15 K There were 2 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_061004 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 2.17 1714.4 1714.8 831.8 0.12 0.14 2.06 2 4.34 3417.6 3417.9 1640.5 -0.20 -0.20 2.08 3 8.67 6841.8 6842.5 3268.8 -0.11 -0.10 2.09 4 13.00 10261.7 10261.7 4977.2 -0.12 -0.12 2.06 5 19.51 15398.9 15398.9 7480.9 -0.08 -0.09 2.06 6 26.01 20544.1 20543.1 10031.5 -0.02 -0.03 2.05 7 32.51 25690.9 25690.2 12256.2 0.02 0.01 2.10 8 43.34 34285.2 34285.8 16471.3 0.11 0.10 2.08 9 54.17 42862.8 42866.3 20148.0 0.12 0.13 2.13 10 65.01 51454.0 51455.0 24113.1 0.16 0.16 2.13 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3654 to 109645 e- Average count rate measured: 790 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.13 +/- 0.01 e-/ADU RON = 4.23 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.004 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.16/-0.20 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay