*** MIDAS logfile (05SEPpl1.2) *** Su, 01 Oct 2006 15:53:41 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Su, 01 Oct 2006 15:53:55 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 061001 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 355.70 stddev = 2.60 FIERA temperature = 273.15 K There were 2 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_061001 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 2.17 1718.0 1718.2 829.0 0.13 0.14 2.07 2 4.34 3426.3 3425.9 1656.6 -0.18 -0.19 2.07 3 8.67 6857.2 6857.1 3261.2 -0.10 -0.10 2.10 4 13.00 10284.4 10284.3 4996.7 -0.12 -0.12 2.06 5 19.51 15433.3 15434.1 7430.8 -0.08 -0.08 2.08 6 26.01 20587.5 20586.8 10171.6 -0.03 -0.04 2.02 7 32.51 25748.0 25747.4 12458.0 0.02 0.02 2.07 8 43.34 34357.6 34358.1 16263.4 0.10 0.10 2.11 9 54.17 42960.2 42961.5 20115.5 0.13 0.13 2.14 10 65.01 51561.2 51562.1 23952.5 0.14 0.15 2.15 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3691 to 110781 e- Average count rate measured: 792 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.15 +/- 0.02 e-/ADU RON = 5.58 +/- 0.05 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.004 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.15/-0.19 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay