*** MIDAS logfile (05SEPpl1.2) *** Fr, 29 Sep 2006 19:48:37 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Fr, 29 Sep 2006 19:48:52 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 060929 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 355.24 stddev = 2.16 FIERA temperature = 273.15 K There were 2 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_060929 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 2.17 1708.5 1708.6 828.5 0.14 0.13 2.06 2 4.34 3408.7 3408.3 1616.3 -0.16 -0.17 2.11 3 8.67 6820.5 6820.9 3297.5 -0.12 -0.11 2.07 4 13.00 10231.6 10231.6 4942.5 -0.12 -0.12 2.07 5 19.51 15354.2 15354.2 7366.8 -0.08 -0.08 2.08 6 26.01 20482.3 20482.7 9941.5 -0.03 -0.03 2.06 7 32.51 25609.9 25611.1 12097.0 -0.00 -0.00 2.12 8 43.34 34184.6 34186.6 16439.5 0.11 0.11 2.08 9 54.17 42727.5 42731.4 19995.1 0.10 0.10 2.14 10 65.01 51307.7 51310.2 24040.6 0.16 0.17 2.13 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3646 to 109494 e- Average count rate measured: 788 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.13 +/- 0.01 e-/ADU RON = 4.61 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.005 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.17/-0.16 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay