*** MIDAS logfile (05SEPpl1.2) *** We, 27 Sep 2006 19:57:12 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** We, 27 Sep 2006 19:57:27 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 060927 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 355.26 stddev = 2.63 FIERA temperature = 273.15 K There were 2 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_060927 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 2.17 1709.7 1709.4 845.8 0.13 0.10 2.02 2 4.34 3409.9 3410.9 1622.2 -0.15 -0.13 2.10 3 8.67 6820.4 6821.2 3258.4 -0.13 -0.13 2.09 4 13.00 10230.3 10237.7 4964.3 -0.14 -0.08 2.06 5 19.51 15352.2 15351.7 7440.6 -0.10 -0.11 2.06 6 26.01 20478.9 20479.5 9865.3 -0.05 -0.06 2.08 7 32.51 25609.9 25610.7 12397.0 -0.01 -0.02 2.07 8 43.34 34201.1 34204.0 16269.6 0.15 0.15 2.10 9 54.17 42730.8 42734.2 20395.2 0.10 0.10 2.10 10 65.01 51330.6 51330.6 24078.2 0.21 0.20 2.13 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3625 to 108834 e- Average count rate measured: 788 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.12 +/- 0.01 e-/ADU RON = 5.58 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.004 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.20/-0.14 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay