*** MIDAS logfile (05SEPpl1.2) *** We, 13 Sep 2006 15:18:06 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** We, 13 Sep 2006 15:18:21 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 060913 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 355.58 stddev = 2.01 FIERA temperature = 273.15 K There were 2 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_060913 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 2.17 1710.2 1710.8 818.1 0.15 0.14 2.09 2 4.34 3409.0 3408.2 1624.7 -0.21 -0.23 2.10 3 8.67 6825.4 6824.9 3263.3 -0.08 -0.10 2.09 4 13.00 10234.8 10235.2 5179.5 -0.12 -0.12 1.98 5 19.51 15360.0 15359.5 7372.1 -0.08 -0.08 2.08 6 26.01 20487.8 20488.3 9943.5 -0.04 -0.03 2.06 7 32.51 25623.6 25624.5 12372.7 0.02 0.03 2.07 8 43.34 34189.8 34190.1 16253.1 0.09 0.10 2.10 9 54.17 42758.3 42758.6 20508.6 0.14 0.15 2.08 10 65.01 51306.5 51307.5 23766.6 0.13 0.14 2.16 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3652 to 109552 e- Average count rate measured: 788 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.14 +/- 0.02 e-/ADU RON = 4.29 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.004 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.15/-0.22 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay