*** MIDAS logfile (05SEPpl1.2) *** We, 06 Sep 2006 14:34:51 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** We, 06 Sep 2006 14:35:06 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 060906 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 355.51 stddev = 2.01 FIERA temperature = 273.15 K There were 2 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_060906 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 2.17 1720.3 1720.1 822.6 0.16 0.13 2.09 2 4.34 3427.7 3428.7 1621.5 -0.22 -0.21 2.11 3 8.67 6863.5 6863.6 3299.6 -0.10 -0.10 2.08 4 13.00 10292.8 10292.6 4989.1 -0.13 -0.13 2.06 5 19.51 15447.1 15447.5 7414.3 -0.08 -0.07 2.08 6 26.01 20605.7 20605.3 10030.4 -0.03 -0.03 2.05 7 32.51 25768.7 25770.3 12473.6 0.01 0.02 2.07 8 43.34 34387.5 34386.7 16355.1 0.10 0.10 2.10 9 54.17 43001.1 43001.2 20426.2 0.14 0.14 2.11 10 65.01 51603.6 51606.1 24073.0 0.14 0.15 2.14 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3663 to 109887 e- Average count rate measured: 793 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.13 +/- 0.02 e-/ADU RON = 4.28 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.004 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.16/-0.22 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay