*** MIDAS logfile (05SEPpl1.2) *** Th, 31 Aug 2006 14:13:35 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Th, 31 Aug 2006 14:13:50 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 060831 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 355.78 stddev = 2.03 FIERA temperature = 273.15 K There were 2 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_060831 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 2.17 1717.9 1718.0 839.0 0.14 0.14 2.05 2 4.34 3424.9 3425.1 1626.5 -0.22 -0.21 2.11 3 8.67 6857.3 6857.5 3272.9 -0.09 -0.09 2.10 4 13.00 10282.6 10282.6 4985.1 -0.13 -0.13 2.06 5 19.51 15432.2 15433.7 7450.6 -0.08 -0.08 2.07 6 26.01 20588.4 20586.1 9996.2 -0.02 -0.04 2.06 7 32.51 25746.1 25748.2 12315.0 0.02 0.02 2.09 8 43.34 34357.1 34355.8 16360.8 0.10 0.10 2.10 9 54.17 42962.3 42964.9 20298.8 0.14 0.14 2.12 10 65.01 51557.6 51559.4 23804.7 0.14 0.15 2.17 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3690 to 110734 e- Average count rate measured: 792 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.15 +/- 0.02 e-/ADU RON = 4.37 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.004 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.14/-0.21 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay