*** MIDAS logfile (05SEPpl1.2) *** Sa, 26 Aug 2006 19:55:55 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Sa, 26 Aug 2006 19:56:10 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 060826 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 355.41 stddev = 2.60 FIERA temperature = 273.15 K There were 2 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_060826 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 2.17 1713.5 1712.6 828.0 0.14 0.13 2.07 2 4.34 3418.4 3418.2 1635.9 -0.14 -0.14 2.09 3 8.67 6835.5 6838.8 3286.6 -0.17 -0.13 2.08 4 13.00 10262.6 10261.9 4861.1 -0.09 -0.11 2.11 5 19.51 15397.8 15398.2 7557.9 -0.08 -0.08 2.04 6 26.01 20542.0 20542.9 9923.7 -0.02 -0.02 2.07 7 32.51 25685.4 25685.8 12371.9 0.01 0.00 2.08 8 43.34 34285.3 34286.4 16568.3 0.12 0.11 2.07 9 54.17 42825.1 42832.3 20370.1 0.05 0.05 2.10 10 65.01 51460.9 51464.7 24047.1 0.19 0.18 2.14 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3634 to 109169 e- Average count rate measured: 790 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.12 +/- 0.02 e-/ADU RON = 5.52 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.006 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.18/-0.15 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay