*** MIDAS logfile (05SEPpl1.2) *** Su, 20 Aug 2006 15:18:45 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Su, 20 Aug 2006 15:18:59 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 060820 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 355.28 stddev = 2.61 FIERA temperature = 273.15 K There were 2 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_060820 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 2.17 1717.6 1718.3 818.6 0.15 0.14 2.10 2 4.34 3423.8 3423.7 1610.7 -0.20 -0.21 2.13 3 8.67 6853.6 6852.9 3284.8 -0.10 -0.12 2.09 4 13.00 10279.1 10278.7 4962.8 -0.12 -0.12 2.07 5 19.51 15423.1 15423.5 7439.6 -0.10 -0.09 2.07 6 26.01 20577.1 20577.7 9841.1 -0.03 -0.02 2.09 7 32.51 25732.4 25731.1 12609.0 0.01 0.01 2.04 8 43.34 34344.6 34342.0 16340.2 0.11 0.11 2.10 9 54.17 42947.8 42948.6 20120.1 0.15 0.16 2.13 10 65.01 51534.9 51531.5 24059.5 0.14 0.15 2.14 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3671 to 110112 e- Average count rate measured: 792 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.14 +/- 0.02 e-/ADU RON = 5.58 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.004 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.16/-0.21 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay