*** MIDAS logfile (05SEPpl1.2) *** Sa, 12 Aug 2006 13:50:07 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Sa, 12 Aug 2006 13:50:21 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 060812 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 356.20 stddev = 2.59 FIERA temperature = 273.15 K There were 2 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_060812 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 2.17 1714.8 1715.3 835.7 0.13 0.13 2.05 2 4.34 3421.1 3420.4 1619.9 -0.17 -0.19 2.11 3 8.67 6846.5 6846.8 3310.4 -0.11 -0.11 2.07 4 13.00 10271.7 10271.5 5010.5 -0.10 -0.11 2.05 5 19.51 15411.2 15412.3 7576.0 -0.09 -0.08 2.03 6 26.01 20559.8 20559.7 9856.4 -0.03 -0.03 2.09 7 32.51 25708.5 25711.3 12319.8 0.00 0.01 2.09 8 43.34 34314.9 34314.8 16204.4 0.11 0.11 2.12 9 54.17 42892.5 42896.5 20078.9 0.11 0.11 2.14 10 65.01 51500.5 51502.7 23760.6 0.16 0.16 2.17 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3707 to 111320 e- Average count rate measured: 791 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.16 +/- 0.02 e-/ADU RON = 5.60 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.005 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.16/-0.18 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay