*** MIDAS logfile (05SEPpl1.2) *** We, 02 Aug 2006 17:21:12 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** We, 02 Aug 2006 17:21:26 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 060802 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 356.01 stddev = 2.60 FIERA temperature = 273.15 K There were 2 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_060802 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1592.0 1591.6 748.8 0.13 0.13 2.13 2 6.16 3173.4 3173.1 1529.3 -0.19 -0.20 2.08 3 12.31 6353.6 6353.3 3014.9 -0.09 -0.10 2.11 4 18.47 9526.6 9526.5 4607.6 -0.14 -0.13 2.07 5 27.70 14299.7 14298.1 7038.8 -0.07 -0.07 2.03 6 36.93 19072.7 19071.1 9266.7 -0.03 -0.04 2.06 7 46.16 23855.5 23855.5 11495.5 0.03 0.03 2.08 8 61.54 31823.5 31822.3 15464.8 0.08 0.08 2.06 9 76.93 39809.6 39807.8 18878.9 0.15 0.15 2.11 10 92.31 47767.2 47764.4 22383.1 0.14 0.14 2.13 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3372 to 101197 e- Average count rate measured: 517 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.12 +/- 0.02 e-/ADU RON = 5.51 +/- 0.05 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.004 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.15/-0.20 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay