*** MIDAS logfile (05SEPpl1.2) *** Th, 20 Jul 2006 16:36:46 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Th, 20 Jul 2006 16:37:01 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 060720 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 355.39 stddev = 2.10 FIERA temperature = 273.15 K There were 2 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_060720 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1593.8 1593.8 762.5 0.13 0.15 2.09 2 6.16 3178.0 3177.4 1502.3 -0.19 -0.20 2.12 3 12.31 6362.0 6362.3 3090.6 -0.10 -0.10 2.06 4 18.47 9540.1 9539.9 4534.0 -0.14 -0.14 2.10 5 27.70 14318.5 14318.4 6933.6 -0.07 -0.07 2.07 6 36.93 19099.1 19098.5 9144.4 -0.04 -0.04 2.09 7 46.16 23888.6 23888.1 11622.2 0.03 0.02 2.06 8 61.54 31869.0 31868.3 15207.9 0.08 0.08 2.10 9 76.93 39866.3 39865.1 18996.3 0.16 0.15 2.10 10 92.31 47835.0 47834.0 22461.6 0.15 0.14 2.13 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3373 to 101246 e- Average count rate measured: 517 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.12 +/- 0.01 e-/ADU RON = 4.44 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.005 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.15/-0.19 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay