*** MIDAS logfile (05SEPpl1.2) *** Su, 02 Jul 2006 14:20:52 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Su, 02 Jul 2006 14:21:06 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 060702 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 355.06 stddev = 2.10 FIERA temperature = 273.15 K There were 2 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_060702 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1590.4 1590.4 751.4 0.13 0.12 2.12 2 6.16 3171.6 3172.0 1510.9 -0.18 -0.18 2.10 3 12.31 6349.1 6349.5 3039.7 -0.11 -0.10 2.09 4 18.47 9521.8 9522.2 4586.9 -0.13 -0.13 2.08 5 27.70 14291.2 14289.7 6900.8 -0.07 -0.08 2.07 6 36.93 19061.7 19061.9 9332.8 -0.04 -0.03 2.04 7 46.16 23839.9 23840.7 11399.7 0.02 0.02 2.09 8 61.54 31807.5 31807.1 15208.5 0.08 0.08 2.09 9 76.93 39784.3 39782.3 18623.0 0.14 0.14 2.14 10 92.31 47745.5 47745.4 22398.2 0.15 0.15 2.13 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3391 to 101804 e- Average count rate measured: 516 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.13 +/- 0.01 e-/ADU RON = 4.48 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.005 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.15/-0.18 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay