*** MIDAS logfile (05SEPpl1.2) *** Sa, 13 May 2006 15:04:56 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Sa, 13 May 2006 15:05:11 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 060513 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 355.65 stddev = 2.06 FIERA temperature = 273.15 K There were 2 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_060513 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1624.6 1624.9 775.2 0.17 0.15 2.10 2 6.16 3236.6 3237.1 1548.7 -0.24 -0.23 2.09 3 12.31 6484.3 6483.6 3131.1 -0.08 -0.08 2.07 4 18.47 9719.0 9718.4 4685.7 -0.15 -0.15 2.07 5 27.70 14591.3 14588.4 7040.1 -0.06 -0.07 2.07 6 36.93 19460.1 19458.4 9340.3 -0.04 -0.03 2.08 7 46.16 24345.0 24342.7 11905.2 0.04 0.05 2.04 8 61.54 32463.8 32461.1 15409.1 0.05 0.06 2.11 9 76.93 40641.1 40634.9 19395.7 0.20 0.20 2.10 10 92.31 48720.3 48714.2 22755.7 0.10 0.10 2.14 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3452 to 103498 e- Average count rate measured: 527 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.12 +/- 0.02 e-/ADU RON = 4.37 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.004 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.20/-0.23 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay