*** MIDAS logfile (05SEPpl1.2) *** Sa, 06 May 2006 15:59:22 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Sa, 06 May 2006 15:59:37 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 060506 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 352.07 stddev = 2.09 FIERA temperature = 273.15 K There were 2 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_060506 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1650.6 1650.3 785.4 0.14 0.14 2.10 2 6.16 3291.0 3290.5 1617.3 -0.19 -0.19 2.03 3 12.31 6588.3 6587.9 3145.6 -0.10 -0.10 2.09 4 18.47 9879.4 9879.0 4807.0 -0.13 -0.13 2.06 5 27.70 14827.8 14827.6 7192.1 -0.07 -0.07 2.06 6 36.93 19778.0 19777.3 9614.6 -0.03 -0.04 2.06 7 46.16 24735.2 24736.7 11960.1 0.02 0.02 2.07 8 61.54 33000.7 32998.7 15699.5 0.08 0.07 2.10 9 76.93 41278.5 41279.9 19569.7 0.15 0.15 2.11 10 92.31 49529.6 49530.3 22961.4 0.14 0.14 2.16 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3534 to 106049 e- Average count rate measured: 536 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.14 +/- 0.02 e-/ADU RON = 4.48 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.005 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.15/-0.19 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay