*** MIDAS logfile (05SEPpl1.2) *** Mo, 01 May 2006 17:01:38 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Mo, 01 May 2006 17:01:52 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 060501 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 351.61 stddev = 2.06 FIERA temperature = 273.15 K There were 2 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_060501 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1646.0 1645.3 775.7 0.13 0.14 2.12 2 6.16 3281.3 3281.8 1556.3 -0.20 -0.17 2.11 3 12.31 6568.8 6568.5 3196.9 -0.11 -0.10 2.05 4 18.47 9851.3 9851.3 4749.9 -0.13 -0.13 2.07 5 27.70 14785.6 14784.8 7042.9 -0.07 -0.07 2.10 6 36.93 19721.3 19720.9 9631.7 -0.03 -0.04 2.05 7 46.16 24665.6 24664.1 11767.3 0.02 0.02 2.10 8 61.54 32908.3 32906.2 15704.4 0.09 0.08 2.10 9 76.93 41156.8 41157.7 19558.5 0.14 0.14 2.10 10 92.31 49393.2 49391.3 22859.4 0.15 0.14 2.16 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3520 to 105641 e- Average count rate measured: 534 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.14 +/- 0.02 e-/ADU RON = 4.40 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.005 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.15/-0.19 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay