*** MIDAS logfile (05SEPpl1.2) *** Fr, 21 Apr 2006 13:56:07 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Fr, 21 Apr 2006 13:56:21 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 060421 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 352.23 stddev = 2.10 FIERA temperature = 273.15 K There were 2 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_060421 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1644.3 1644.4 789.6 0.12 0.13 2.08 2 6.16 3281.8 3282.9 1557.9 -0.14 -0.12 2.11 3 12.31 6566.8 6567.4 3121.3 -0.12 -0.13 2.10 4 18.47 9852.4 9852.5 4818.3 -0.11 -0.12 2.04 5 27.70 14783.8 14784.7 7184.4 -0.08 -0.08 2.06 6 36.93 19721.6 19723.7 9555.0 -0.04 -0.03 2.06 7 46.16 24661.5 24661.9 12057.9 -0.00 -0.01 2.05 8 61.54 32913.5 32914.4 15780.4 0.09 0.09 2.09 9 76.93 41147.4 41151.7 19667.8 0.11 0.11 2.09 10 92.31 49411.4 49415.2 23013.4 0.17 0.17 2.15 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3492 to 104950 e- Average count rate measured: 534 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.12 +/- 0.02 e-/ADU RON = 4.46 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.007 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.17/-0.13 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay