*** MIDAS logfile (05SEPpl1.2) *** Fr, 14 Apr 2006 16:50:26 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Fr, 14 Apr 2006 16:50:41 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 060414 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 353.01 stddev = 2.10 FIERA temperature = 273.15 K There were 2 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_060414 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1652.6 1651.9 795.7 0.13 0.13 2.08 2 6.16 3295.4 3295.2 1569.6 -0.18 -0.18 2.10 3 12.31 6596.5 6595.9 3136.5 -0.11 -0.12 2.10 4 18.47 9893.0 9893.4 4801.5 -0.13 -0.13 2.06 5 27.70 14848.8 14847.9 7347.0 -0.07 -0.08 2.02 6 36.93 19806.7 19806.9 9551.7 -0.03 -0.03 2.07 7 46.16 24770.4 24771.0 12142.7 0.02 0.02 2.04 8 61.54 33051.6 33052.8 15875.5 0.09 0.09 2.08 9 76.93 41331.5 41334.4 19570.0 0.13 0.14 2.11 10 92.31 49610.7 49613.8 23225.3 0.16 0.16 2.14 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3510 to 105406 e- Average count rate measured: 537 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.12 +/- 0.02 e-/ADU RON = 4.45 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.006 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.16/-0.18 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay