*** MIDAS logfile (05SEPpl1.2) *** Fr, 07 Apr 2006 14:13:18 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Fr, 07 Apr 2006 14:13:33 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 060407 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 352.57 stddev = 2.09 FIERA temperature = 273.15 K There were 2 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_060407 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1647.5 1648.0 789.1 0.11 0.13 2.09 2 6.16 3287.9 3287.4 1569.0 -0.16 -0.18 2.10 3 12.31 6580.1 6580.4 3150.4 -0.11 -0.11 2.09 4 18.47 9870.0 9869.9 4800.4 -0.12 -0.12 2.06 5 27.70 14811.1 14811.6 7219.7 -0.08 -0.08 2.05 6 36.93 19757.2 19758.7 9454.8 -0.04 -0.03 2.09 7 46.16 24710.1 24709.7 11890.6 0.02 0.01 2.08 8 61.54 32971.9 32971.2 16088.0 0.09 0.09 2.05 9 76.93 41230.9 41234.4 19533.0 0.13 0.14 2.11 10 92.31 49494.1 49494.5 23141.3 0.16 0.16 2.14 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3495 to 104985 e- Average count rate measured: 535 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.12 +/- 0.02 e-/ADU RON = 4.43 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.006 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.16/-0.17 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay