*** MIDAS logfile (05SEPpl1.2) *** Fr, 17 Mar 2006 18:53:32 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Fr, 17 Mar 2006 18:53:51 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 060317 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 352.04 stddev = 2.09 FIERA temperature = 273.15 K There were 2 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_060317 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1677.0 1677.4 809.9 0.12 0.12 2.07 2 6.16 3345.5 3346.0 1604.7 -0.17 -0.16 2.08 3 12.31 6696.2 6695.9 3220.6 -0.11 -0.11 2.08 4 18.47 10042.5 10042.9 4839.0 -0.13 -0.13 2.08 5 27.70 15072.3 15071.6 7280.7 -0.07 -0.08 2.07 6 36.93 20105.1 20105.7 9725.7 -0.03 -0.03 2.07 7 46.16 25142.6 25142.4 12067.2 0.01 0.01 2.08 8 61.54 33551.0 33550.4 16044.3 0.10 0.09 2.09 9 76.93 41948.5 41952.2 19966.3 0.12 0.13 2.10 10 92.31 50357.7 50360.8 23721.9 0.16 0.16 2.12 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3548 to 106532 e- Average count rate measured: 545 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.12 +/- 0.01 e-/ADU RON = 4.43 +/- 0.02 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.005 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.16/-0.17 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay