*** MIDAS logfile (05SEPpl1.2) *** Fr, 03 Mar 2006 16:22:40 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Fr, 03 Mar 2006 16:22:54 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 060303 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 355.58 stddev = 2.05 FIERA temperature = 273.15 K There were 2 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_060303 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1676.1 1675.9 812.9 0.18 0.15 2.06 2 6.16 3337.3 3337.8 1591.7 -0.28 -0.26 2.10 3 12.31 6689.1 6688.6 3196.1 -0.07 -0.07 2.09 4 18.47 10025.1 10024.1 4799.5 -0.15 -0.15 2.09 5 27.70 15049.5 15047.8 7330.2 -0.07 -0.07 2.05 6 36.93 20072.8 20070.1 9742.1 -0.04 -0.04 2.06 7 46.16 25114.1 25111.1 11997.0 0.05 0.05 2.09 8 61.54 33483.7 33481.6 15910.6 0.05 0.06 2.10 9 76.93 41934.3 41924.7 19732.6 0.24 0.23 2.13 10 92.31 50246.9 50244.4 23265.6 0.09 0.10 2.16 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3599 to 107907 e- Average count rate measured: 544 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.15 +/- 0.02 e-/ADU RON = 4.41 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.004 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.23/-0.27 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay