*** MIDAS logfile (05SEPpl1.2) *** Sa, 25 Feb 2006 13:02:07 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Sa, 25 Feb 2006 13:02:22 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 060225 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 355.54 stddev = 2.05 FIERA temperature = 273.15 K There were 2 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_060225 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1688.8 1689.1 813.6 0.13 0.12 2.08 2 6.16 3369.3 3369.4 1599.6 -0.16 -0.18 2.11 3 12.31 6744.3 6744.3 3215.7 -0.10 -0.10 2.10 4 18.47 10114.6 10114.6 4802.0 -0.13 -0.13 2.11 5 27.70 15179.7 15178.9 7354.7 -0.08 -0.08 2.06 6 36.93 20248.5 20248.3 9836.5 -0.04 -0.03 2.06 7 46.16 25323.5 25323.5 12232.7 0.01 0.02 2.07 8 61.54 33788.6 33788.7 16123.8 0.08 0.09 2.10 9 76.93 42257.9 42258.9 19850.8 0.14 0.14 2.13 10 92.31 50715.2 50713.7 23527.9 0.15 0.15 2.16 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3618 to 108640 e- Average count rate measured: 549 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.14 +/- 0.02 e-/ADU RON = 4.39 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.006 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.15/-0.17 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay