*** MIDAS logfile (05SEPpl1.2) *** Fr, 17 Feb 2006 14:57:01 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Fr, 17 Feb 2006 14:57:16 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 060217 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 355.43 stddev = 2.07 FIERA temperature = 273.15 K There were 2 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_060217 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1687.4 1687.7 816.8 0.11 0.11 2.07 2 6.16 3368.0 3368.4 1597.5 -0.15 -0.15 2.11 3 12.31 6741.4 6740.7 3209.8 -0.11 -0.12 2.10 4 18.47 10111.6 10112.7 4845.6 -0.12 -0.11 2.09 5 27.70 15174.1 15173.4 7459.3 -0.08 -0.09 2.03 6 36.93 20242.4 20244.3 9764.5 -0.03 -0.03 2.07 7 46.16 25314.5 25314.1 12106.5 0.01 0.00 2.09 8 61.54 33781.7 33782.0 16147.6 0.09 0.09 2.09 9 76.93 42238.9 42241.7 19925.2 0.12 0.12 2.12 10 92.31 50708.9 50710.1 23475.0 0.16 0.16 2.16 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3618 to 108717 e- Average count rate measured: 548 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.14 +/- 0.02 e-/ADU RON = 4.43 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.007 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.16/-0.15 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay