*** MIDAS logfile (05SEPpl1.2) *** Sa, 11 Feb 2006 14:01:52 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Sa, 11 Feb 2006 14:02:07 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 060211 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 355.22 stddev = 2.06 FIERA temperature = 273.15 K There were 2 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_060211 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1703.4 1703.2 813.4 0.12 0.14 2.09 2 6.16 3397.9 3397.6 1614.7 -0.17 -0.17 2.10 3 12.31 6800.6 6800.9 3236.8 -0.11 -0.11 2.10 4 18.47 10199.9 10200.1 4916.3 -0.13 -0.12 2.07 5 27.70 15307.9 15308.0 7367.1 -0.07 -0.08 2.08 6 36.93 20419.7 20419.2 9779.1 -0.03 -0.04 2.09 7 46.16 25537.2 25537.0 12351.9 0.02 0.01 2.07 8 61.54 34074.9 34075.3 16419.5 0.09 0.09 2.08 9 76.93 42606.2 42609.3 20258.8 0.12 0.12 2.10 10 92.31 51144.7 51147.4 23875.3 0.15 0.15 2.14 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3614 to 108530 e- Average count rate measured: 553 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.12 +/- 0.02 e-/ADU RON = 4.38 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.006 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.15/-0.17 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay