*** MIDAS logfile (05SEPpl1.2) *** Su, 05 Feb 2006 16:34:49 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Su, 05 Feb 2006 16:35:04 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 060205 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 355.43 stddev = 2.06 FIERA temperature = 273.15 K There were 2 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_060205 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1695.9 1695.4 822.0 0.13 0.13 2.06 2 6.16 3383.0 3383.0 1630.0 -0.17 -0.16 2.08 3 12.31 6770.8 6770.3 3213.0 -0.11 -0.12 2.11 4 18.47 10156.1 10155.8 4872.7 -0.12 -0.13 2.08 5 27.70 15241.2 15242.5 7391.0 -0.08 -0.07 2.06 6 36.93 20331.8 20332.3 9770.1 -0.03 -0.03 2.08 7 46.16 25425.7 25426.6 12237.1 0.01 0.01 2.08 8 61.54 33928.3 33928.7 16013.4 0.09 0.09 2.12 9 76.93 42423.6 42424.8 20060.9 0.12 0.12 2.11 10 92.31 50923.3 50927.2 23847.3 0.15 0.15 2.14 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3614 to 108537 e- Average count rate measured: 551 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.13 +/- 0.01 e-/ADU RON = 4.39 +/- 0.02 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.007 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.15/-0.16 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay