*** MIDAS logfile (05SEPpl1.2) *** Fr, 27 Jan 2006 13:11:24 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Fr, 27 Jan 2006 13:11:38 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 060127 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 354.51 stddev = 2.07 FIERA temperature = 273.15 K There were 2 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_060127 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1697.4 1697.8 818.6 0.12 0.13 2.07 2 6.16 3386.7 3386.7 1609.2 -0.17 -0.17 2.10 3 12.31 6778.2 6777.7 3242.3 -0.11 -0.12 2.09 4 18.47 10166.5 10166.6 4921.5 -0.13 -0.13 2.07 5 27.70 15257.3 15258.3 7314.8 -0.08 -0.07 2.09 6 36.93 20354.6 20354.1 9951.0 -0.03 -0.03 2.05 7 46.16 25454.9 25453.0 12405.9 0.02 0.01 2.05 8 61.54 33963.7 33964.2 16303.1 0.09 0.09 2.08 9 76.93 42475.2 42475.0 19953.8 0.14 0.13 2.13 10 92.31 50979.1 50981.3 23745.5 0.15 0.15 2.15 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3625 to 108870 e- Average count rate measured: 551 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.14 +/- 0.02 e-/ADU RON = 4.43 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.006 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.15/-0.17 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay