*** MIDAS logfile (05SEPpl1.2) *** Sa, 14 Jan 2006 15:22:24 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Sa, 14 Jan 2006 15:22:40 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 060114 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 355.71 stddev = 2.06 FIERA temperature = 273.15 K There were 2 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_060114 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1694.7 1694.4 836.6 0.14 0.13 2.03 2 6.16 3381.4 3381.2 1625.5 -0.16 -0.16 2.08 3 12.31 6767.9 6767.9 3230.2 -0.12 -0.11 2.10 4 18.47 10151.6 10150.9 4865.9 -0.13 -0.13 2.09 5 27.70 15234.6 15235.6 7413.4 -0.08 -0.07 2.06 6 36.93 20323.9 20322.1 9799.4 -0.03 -0.04 2.07 7 46.16 25417.5 25416.1 12282.5 0.01 0.01 2.07 8 61.54 33915.8 33917.0 16205.5 0.09 0.10 2.09 9 76.93 42408.4 42412.6 19997.7 0.12 0.13 2.12 10 92.31 50908.5 50909.9 23896.6 0.16 0.16 2.13 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3604 to 108266 e- Average count rate measured: 551 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.13 +/- 0.01 e-/ADU RON = 4.37 +/- 0.02 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.007 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.16/-0.16 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay