*** MIDAS logfile (05SEPpl1.2) *** We, 11 Jan 2006 15:45:11 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** We, 11 Jan 2006 15:45:25 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 060111 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 355.76 stddev = 2.04 FIERA temperature = 273.15 K There were 2 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_060111 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1702.2 1702.2 821.3 0.13 0.15 2.07 2 6.16 3398.4 3398.5 1627.9 -0.13 -0.14 2.09 3 12.31 6800.4 6800.1 3230.6 -0.12 -0.15 2.10 4 18.47 10202.0 10201.4 4956.0 -0.12 -0.15 2.06 5 27.70 15308.2 15308.9 7475.5 -0.09 -0.12 2.05 6 36.93 20422.6 20424.3 9771.1 -0.03 -0.06 2.09 7 46.16 25537.3 25537.4 12358.2 -0.00 -0.04 2.07 8 61.54 34084.1 34084.8 16355.7 0.10 0.06 2.08 9 76.93 42600.0 42609.0 19994.9 0.08 0.06 2.13 10 92.31 51169.3 51299.6 23958.1 0.18 0.39 2.14 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3632 to 109308 e- Average count rate measured: 553 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.13 +/- 0.02 e-/ADU RON = 4.36 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.009 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.29/-0.14 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay