*** MIDAS logfile (05SEPpl1.2) *** Th, 29 Dec 2005 14:34:11 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Th, 29 Dec 2005 14:34:26 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 051229 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 356.81 stddev = 1.99 FIERA temperature = 273.15 K There were 2 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_051229 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1709.2 1709.2 827.9 0.13 0.13 2.06 2 6.16 3410.0 3409.8 1639.5 -0.17 -0.17 2.08 3 12.31 6825.0 6824.8 3222.8 -0.12 -0.12 2.12 4 18.47 10237.2 10237.2 4955.5 -0.13 -0.13 2.07 5 27.70 15364.4 15363.2 7449.4 -0.07 -0.08 2.06 6 36.93 20495.1 20494.1 9903.7 -0.03 -0.04 2.07 7 46.16 25631.4 25630.9 12252.7 0.02 0.01 2.09 8 61.54 34202.8 34203.9 16197.7 0.10 0.10 2.11 9 76.93 42762.7 42769.0 20355.6 0.12 0.13 2.10 10 92.31 51339.8 51345.0 23985.3 0.16 0.17 2.14 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3639 to 109313 e- Average count rate measured: 555 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.13 +/- 0.01 e-/ADU RON = 4.24 +/- 0.02 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.006 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.17/-0.17 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay